Product description

PSx series, AC/AC group, like other displacement transducers made by Peltron are based on the differential transformer principle – LVDT.  The transformer is placed inside a cylindrical case. There is a magnetic core inside the transformer. Displacement of the core results in a change in the transducer’s output signal.  Because this series is equipped with a return spring so the core does not has to be fixed to a moveable object. One transducer of  PSx series works with one channel conditioner like amplifier WG06.  These transducers are used in static and dynamic measurements like: displacements, materials length and thickness change, construction and machine elements deflections.

Documents for download

Technical parameters

Measure ranges [mm] 1,2,6,10,20,30,50,100, 150, 200
Power supply 2 Vrms, 5÷15 mA, 5 kHz
Output signal ±1 Vrms ±10%
Basic accuracy [%] MR ▪ ≤0,5
▪ ≤0,25
▪ ≤0,1
Operating temperature [°C] ▪ -20÷+80 standard
▪ -20÷+120
▪ -20÷+200
Ø body 14 mm (19mm)
Ø core 4  mm (5mm)
Electrical connection

▪ PVC cable ½ m, no connector - standard

▪ PVC cable ½ m plus connector - option

▪ connector attached to body

Case material 316L
Return spring Yes